Skip to main content

Inspection visit

Fire Safety inspection

SHERWOOD OAKS POST ACUTE CARE, LLCCMS #05648315 citations on this visit
15 citations recorded

Citations

15 citations recorded*CMS

This is a Fire Safety survey. CMS cites these findings under the Life Safety Code and Emergency Preparedness rules, a separate set from the health (F-tag) deficiencies. The codes and grading differ, so these do not link to the health violation pages.

What do CMS severity letters mean?

Serious (G-L). Actual harm to a resident, or immediate jeopardy. Codes G through I indicate actual harm; J through L indicate immediate jeopardy to resident health or safety.

General (A-F). No actual harm found, or harm that is minimal. The facility must still submit a Plan of Correction. Most CMS citations land here.

Each letter combines severity with scope: how many residents the deficiency affected.

  • 0321GeneralS&S Dpotential for harm

    Ensure that special areas are constructed so that walls can resist fire for one hour or have an approved fire extinguishing system.

  • 0324GeneralS&S Dpotential for harm

    Provide properly protected cooking facilities.

  • 0345GeneralS&S Dpotential for harm

    Have approved installation, maintenance and testing program for fire alarm systems.

  • 0347GeneralS&S Fpotential for harm

    Properly provide smoke detection systems in areas open to corridors.

  • 0353GeneralS&S Fpotential for harm

    Inspect, test, and maintain automatic sprinkler systems.

  • 0363GeneralS&S Dpotential for harm

    Install corridor and hallway doors that block smoke.

  • 0712GeneralS&S Fpotential for harm

    Have simulated fire drills held at unexpected times.

  • 0741GeneralS&S Dpotential for harm

    Have posted "No-smoking" signs in areas where smoking is not permitted or ashtrays provided where smoking was allowed.

  • 0781GeneralS&S Dpotential for harm

    Have restrictions on the use of portable space heaters.

  • 0914GeneralS&S Fpotential for harm

    Ensure receptacles at patient bed locations and where general anesthesia is administered, are tested after initial installation, replacement or servicing.

  • 0918GeneralS&S Fpotential for harm

    Have generator or other power source capable of supplying service within 10 seconds.

  • 0919GeneralS&S Dpotential for harm

    Meet requirements for the use of electrical equipment.

  • 0920GeneralS&S Dpotential for harm

    Ensure proper usage of power strips and extension cords.

  • 0926GeneralS&S Fpotential for harm

    Ensure that personnel concerned with handling of medical gases and cylinders are trained on the risk.

  • 0351GeneralS&S Dpotential for harm

    Install an approved automatic sprinkler system.

FAQ · About this visit

Common questions about this visit

What happened during the May 24, 2024 survey of SHERWOOD OAKS POST ACUTE CARE, LLC?

This was a inspection survey of SHERWOOD OAKS POST ACUTE CARE, LLC on May 24, 2024. The surveyor cited 15 deficiencies, recorded on the federal Form 2567 statement of deficiencies.

Were any deficiencies cited at SHERWOOD OAKS POST ACUTE CARE, LLC on May 24, 2024?

Yes, 15 deficiencies were cited, each with a CMS Scope and Severity grade. The first was: "Ensure that special areas are constructed so that walls can resist fire for one hour or have an approved fire extinguish..."

What type of survey was this?

This was a inspection survey conducted by state surveyors under federal Centers for Medicare & Medicaid Services (CMS) oversight. Findings are published on CMS Care Compare.

Share this reportEmail

Next steps

Concerned about a resident’s care?Find your local ombudsman through the Eldercare Locatoror file a complaint with your state survey agency.

Researching this visit professionally?Book a 15-minute calland we will walk through what we have on file.