F 0880
Provide and implement an infection prevention and control program.
Level of Harm - Minimal harm
or potential for actual harm
Based observation, interview, and record review the facility failed to implement infection control practices for
three of four emergency crash carts when:
Residents Affected - Few
1. Emergency crash carts (crash cart is a set of trays/drawers/shelves on wheels used for transportation
and dispensing of emergency medication/equipment at site of medical/surgical emergency for life support
protocols to potentially save someone's life) B, and C contained oropharyngeal airway (OPA, also known as
an oral airway is a medical device that helps maintain or keep a patient's airway open) kit that was not
stored in the original packaging and was not labeled with shelf-life expectancy or no expiration date.
2. Emergency crash cart A contained oropharyngeal airway kit stored in two plastic bags had yellowish
substance with no label or expiration date.
3. Emergency crash care B and C had one yankauer (Yankauer suction tips are often used to suction
oropharyngeal secretions to prevent aspiration. They can be used in the mouth to remove excess fluid and
secretions from people who have difficulty swallowing or spitting) and one adult nasal oxygen cannula
(nasal cannula is a device that gives you additional oxygen- supplemental oxygen or oxygen therapy
through your nose and It's a thin, flexible tube that goes around your head and into your nose) with no
shelf-life expectancy or expiration date.
These failures had the potential to spread infection and cross-contamination in the facility.
Findings:
1. During an observation with the director of nursing on 7/12/24 at 12:08 p.m., emergency crash cart b and
c contained oropharyngeal airway with no label or expiration date and was not stored in their original
packaging.
During a concurrent observation and interview with the infection control preventionist nurse (ICPN) on
7/12/24 at 12:20 p.m., the ICPN confirmed emergency crash cart B and C contained OPA with different
sizes stored in two separate plastic bags. The first plastic bag had five pieces of OPA 50 millimeter (mm, is
a unit of length in the International System of Units, equal to one thousandth of a meter), five pieces of OPA
60 mm, 70 mm, and 80 mm. The second bag had five pieces of 90 mm, 100 mm, and 110 mm. The ICPN
confirmed the OPAs were not stored in their original packaging and the plastic bags did not have the OPAs
shelf-life expectancy or expiration date. The ICPN further stated that the OPAs should have been stored in
their original packaging for staff to know when the expiration date or shelf-life expectancy.
(continued on next page)
Any deficiency statement ending with an asterisk (*) denotes a deficiency which the institution may be excused from correcting providing it is determined that other
safeguards provide sufficient protection to the patients. (See instructions.) Except for nursing homes, the findings stated above are disclosable 90 days following the
date of survey whether or not a plan of correction is provided. For nursing homes, the above findings and plans of correction are disclosable 14 days following the date
these documents are made available to the facility. If deficiencies are cited, an approved plan of correction is requisite to continued program participation.
LABORATORY DIRECTOR'S OR PROVIDER/SUPPLIER
REPRESENTATIVE'S SIGNATURE
TITLE
(X6) DATE
FORM CMS-2567 (02/99)
Previous Versions Obsolete
Facility ID:
If continuation sheet
Page 1 of 2
Event ID:
555068
Printed: 05/15/2026
Form Approved OMB
No. 0938-0391
Department of Health & Human Services
Centers for Medicare & Medicaid Services
STATEMENT OF DEFICIENCIES
AND PLAN OF CORRECTION
(X1) PROVIDER/SUPPLIER/CLIA
IDENTIFICATION NUMBER:
(X2) MULTIPLE CONSTRUCTION
555068
B. Wing
A. Building
(X3) DATE SURVEY
COMPLETED
08/19/2024
NAME OF PROVIDER OR SUPPLIER
STREET ADDRESS, CITY, STATE, ZIP CODE
White Blossom Care Center
1990 Fruitdale Avenue
San Jose, CA 95128
For information on the nursing home's plan to correct this deficiency, please contact the nursing home or the state survey agency.
(X4) ID PREFIX TAG
SUMMARY STATEMENT OF DEFICIENCIES
(Each deficiency must be preceded by full regulatory or LSC identifying information)
F 0880
Level of Harm - Minimal harm
or potential for actual harm
During a concurrent observation and interview with the central supply (CS) on 7/12/24 at 12:16 p.m., he
acknowledged the above observation. The CS stated that he was responsible for re-stocking medical
supplies in the crash cart, but the above OPAs were there for some time. The CS confirmed the OPAs were
stored in the plastic bags and not from their original packaging. He further stated that he tried to look for the
original packages inside the central supply room but could not find it.
Residents Affected - Few
2. During a concurrent observation and interview with the ICPN on 7/12/24 at 12:28 pm., she confirmed
that emergency crash cart A contained OPAs with different sizes were stored in plastic bags. The first
plastic bag had five pieces of OPA 50 mm, 60 mm, 70 mm, and 80 mm. The second bag had five pieces of
OPA 90 mm, 100 mm, and 110 mm. The ICPN confirmed there was a yellowish substance in the two plastic
bags containing OPAs and the OPAs were not labeled with their shelf-life expectancy or their expiration
date. The ICPN stated the two bags should have been replaced by the staff and the OPAs should have
been stored from their original packaging, for facility staff to now the OPAs shelf-life expectancy or
expiration date.
3. During a concurrent observation and interview with the ICPN on 7/12/24 at 12:30 p.m., the ICPN
confirmed that emergency crash cart for station B and C contained one yankauer with bulb tip and one
adult nasal cannula oxygen tubing with no shelf-life expectancy or no expiration date. The ICPN stated the
yankauer and the nasal cannula oxygen tubing should have shelf-life expectancy or expiration date.
Review of the facility's policy and procedure revised January 2023 titled, Emergency Cart (E-Cart) indicated
to organize and maintain the emergency cart (E- Cart) to ensure adequate needed equipment E- Carts will
be maintained and supplied with required check list which includes respiratory equipment All nurses should
be familiar with the E- Cart contents, location, supplies shall be re -informed during designated mandatory
in services.
Review of the facility's Emergency Cart Check off list, indicated Central Supply will check and refill carts
during weekdays, and supervisor /nurses will check and log on weekends.
Review of the facility's policy and procedure revised February 2023 titled, Medication labeling and Storage
indicated medical supplies and equipment that do not have expiration dates the facility can replace those
items as needed . (e.g.: if /when the packaging or equipment are wet or discolored, visibly worn out.
FORM CMS-2567 (02/99)
Previous Versions Obsolete
Event ID:
Facility ID:
555068
If continuation sheet
Page 2 of 2